Products

Bendtsen

Product line: S624000

For the determination of surface roughness and air permeance according to Bendtsen method. The Bendtsen measurement process is specified both for surface roughness and air permeability. It works according to the air leak method with preset pressure differentials. The ModularLine Bendtsen testing device is available in various different variant models. According to configuration the roughness of the upper surface and / or the lower surface as well as the air permeability can be measured. A device can therefore contain up to three measurement units, each equipped with a closed measurement and regulation system and the latest generation of sensors. Barometric pressure compensation gives you the most accurate results and reproducibility. The results can be displayed in ml/min or μm/Pa*s as required. Calculation according to Gurley is also possible.

Test Description
After inserting the sample and the start of the test sequence, the roughness or air permeance measurement head is automatically placed on the sample and measurement is started. Automated positioning of the head almost rules out inaccuracies of measurement due to outside influence. Depending on the selected measurement pressure a pressure difference of 0.74, 1.47 or 2.20 kPa is created and the throughflow between measurement blade and sample is measured. The air permeability is determined using the same technique, where the airflow through the sample is measured. The same selectable pressure regulation bands apply here.
Applicable Standards
  • ISO 8791-2 roughness / smoothness / ISO 5636-1/-3 air permeance / DIN 53108 roughness (replaced)
    *more standards available
Features
  • Easy operation via inbuilt touch screen
  • Automatic and manual measuring process
  • Pressure difference adjustable
  • Simultaneous measurement of air permeance and roughness
  • Compensation of the barometric pressure
Specification
  • Ease of use due to big touch screen
  • Separate start and stop buttons
  • Compensation of barometric air pressure
  • Measuring range 25-5000 ml/min standard (depending on sensor)
  • Measurement time: 10s (after pressure is reached)
  • Difference pressure adjustable: 0.74 kPa, 1.47 kPa, 2.20 kPa
  • Compatible ProbNet
  • Can be used as ModularLine Module
  • Protection class: 2
  • Protection degree: IP44
Available Models
  • S624000001 Bendtsen 1 head – porosity
  • S624000002 Bendtsen 1 head – roughness top side (standard)
  • S624000003 Bendtsen 1 head – roughness bottom side
  • S624000004 Bendtsen 2 head– roughness b. side / t. side
  • S624000005 Bendtsen 2 head– roughness t. side / porosity
  • S624000006 Bendtsen 2 head– roughness b. side / porosity
  • S624000007 Bendtsen 3 head– roughness b. & t. side / poro. (Special construction with wider housing)
    * more versions and sensors on request
Available Sensors
  • See data sheet
Available Accessories
  • S406900001 ProbeNet software incl. 1 licence
    * requirements explained upon request
Technical data
  • Power supply: 100-240V 50/60Hz
  • Power consumption: <80 W
  • Compressed air connection: 6 bar / hose 6mm
  • Water connection: no
Data
  • Ethernet data output / MQTT
  • USB Updates / Service
  • RS232 data output
Measurement           
  • Dwelling pressure: Dead weight of the measuring head roughness 267g
  • Measuring range: Depending on sensor: 25-5000 standard
  • Measuring time: From reaching the measuring pressure: 10 s
  • Selectable measuring pressure: 0,74 / 1,47 / 2,20 kPa
  • Measuring accuracy: Depending on sensor
  • Pressure difference: +-0,005kPa linear
  • Flow rate: +-0,5% of final value
  • Roughness measuring blade: Ø 31,5mm 150 µm
  • Porosity measuring area: 10cm² +-0,2cm²
  • Results:Air permeability P [µm/Ps] / Air permeability V [ml/min] / Roughness [ml/min] / Gurley [s] (Calculated)
  • Statistics: Mean values / Standard deviation / Coefficient of variation / Min. and max. of test series
Device Dimensions
  • S624000001-6 / Version 1-2 heads / 600 x 600 x 270 mm (L x B x H)
  • S624000007 / Version 3 heads / 600 x 600 x 330 mm (L x B x H)
Weight/ Version / netto / brutto
  • S624000001-6 / Version 1-2 heads / approx. 45kg / approx. 72kg
  • S624000007 / Version 3 heads / approx. 51kg / approx. 77kg

Privacy settings

You can adjust the settings for the cookies here.

Please select which cookies you would like to accept and confirm by clicking the button. You will then be redirected back to the start page.

*Please note that if you check the External media (Google Maps and YouTube) box, data such as your IP address bay be transmitted to Google and thus possibly to a third country without a data protection agreement. This happens especially when you use Google Maps on the site or watch our videos at the products. You can find more information here: Privacy Policy

Country representations, regarding contacts and videos are displayed only if you have checked External media.