Spectro Analyzer
Product line: S40606
For determination of ISO brightness, color, color differences, fluorescence and opacity.The Spectro Analyzer perfectly meets requirements for fast and accurate measurement of optical characteristics of paper during the manufacturing process. The optical system, using a d:0° sphere geometry with 30mm measuring aperture, is in full compliance with all relevant international standards such as ISO, SCAN, DIN, TAPPI, CPPA and AFNOR. Using the patented Numerical UV control (NUVC) it is a highly accurate and reliable bench top spectrophotometer in the production environment. Controlled via tablet panel and the special software the Spectro Analyzer can give values for color, tint, whiteness, yellowness, brightness, fluorescence and opacity. Furthermore, the numerical UV-control allows easy adjustment of the instrument's parameters to other instruments and thus, guarantees perfect data correlation.
Test description
The desired measurement method is selected from preset standard test types or a predefined test program created by the operator. The identification number of the sample (tambour number, etc.) is entered to identify the sample. Then the sample is placed on the sample support, and this is released to initiate automatic closing. Pushing the start button begins the measurement. The results are displayed on the touch screen, both numerically and graphically. If more than one test is carried out, these can be compared as statistics as well as displayed as standard deviations. The data can be easily printed via the unit’s USB port, or stored using a USB compatible storage device.
Applicable standards
- DIN 53145 53147, 54500
- ISO 2469, 2470, 2471, 3688, 11475, 11476
- TAPPI T519, T525, T527
Features
- Compact device with inbuilt touch screen
- Double-beam-spectrometer with d/0° geometry
- Measuring with different light sources without recalibration
- Automatic calculation of the standard deviation after several test sequences
Specification
- Optical system with d:0° spherical geometry complies with all common standards
- Easy adaptation of instrument parameters to other instruments
- Good compatibility with predecessor and its measurement data
- Improved target acquisition through camera preview
- Smaller measuring range possible
- ProbeNet compatible
- Modularline version available (coming soon)
- Protection classification IP20 / 2
System
- Sensor: silicon photodiode cell (2x40 elements) diffraction grating
- Light source: Xenon flash lamps (3 pieces)
- Device comparability: mean ΔE* 0,20 12 BCRA Series II (ceramic tiles compared to the master device under KM calibration conditions)
- UV balance: 100% / 0% / adjusted
- 400 nm and 420 nm UVblocking filters (NUVC: numerical UV adjustment, no mechanical filter movement required)
- Sample viewing: means of built-in camera; software (not included)
System
- Silizium-Photodiodenzelle (2x40 Elemente) Beugungsgitter
- Lichtquelle: Xenon Blitzlampen (3Stück)
- Geräteübereinstimmung: Mittelwert ΔE* 0,20 12 BCRA Serie II (Keramikkacheln verglichen mit Mastergerät unter KM- Kalibrierbedingungen)
- UV Abgleich: 100% / 0% / Angepasst 400 nm und 420 nm UV- Sperrfilter (NUVC: numerischer UV- Abgleich, keine mechanische Filterbewegung erforderlich)
- Probenbetrachtung: Mittels eingebauter Kamera
Available Models
- S406060000 Spectro Analyzer
Available Accessories
- S406051009 UV tile (working standard)
(The UV tile, if not owned already, is recommended for the new purchase of the device)
- S406900001 ProbeNet software inclusive one licence
*requirements gladly explained on request
Technical data
- Power supply: 100-240V 50/60Hz
Power consumption: <35W / 0,04 kWh
Water connection: no
Compressed air connection: no
Data
- USB Updates / Service
- Ethernet data output / MQTT
Measurement
- Measurement method:
- ISO whiteness (R457)
- Color: XYZ/Rx,Ry,Rz/L,a,b/ L*,a*,b*/L*,C*,h*/ x,y,Y
- Color difference between two samples / standard and sample
- Fluorescence measurement
- Opacity
- Measurement geometry: Reflection: d:0° (diffuse lighting / = ° observation)
LAV setup corresponds to the standards ISO 2469, JIS P8148, DIN 53145-1, DIN 53145-2
- Wavelength range: 360 nm -740nm
- Wavelength interval: 10 nm
- Measuring range: 0 – 200%; Resolution: 0,01%
- Measurement time: approx. 1,5 secs
- Minimum time between two measurements: Approx. 2 seconds (UV 100%)Approx. 3 seconds (UV 0% UV adjusted)
- Measurement / lighting area:
LAV : Ø 30/34mm
MAV: Ø 08/11mm
Device dimensions
- S406060000 Spectro Analyzer 340 x 620 x 630 mm (L x B x H)
- screen 300x245
Weight / Version / net / gross
- S406060000 / Spectro Analyzer / approx. 20,5 kg / approx. 60 kg
- with case / approx. 21,5 kg / approx. 60 kg